Conductivity Analyzers

ABB conductivity analysers are designed to meet the most exacting requirements of our valued customers. The extensive range covers 2 and 4-wire analysers and single or dual-input devices.

Our hazardous area certifications embrace FM, CSA and CENELEC ATEX intrinsic safety. Instruments are available with HART, Profibus PA or DP, and Fieldbus Foundation communications capabilities.

2-Electrode Contacting Cells - Range 0 to 10,000 µS/cm
High accuracy technique for trace conductivities from ultrapure water to industrial rinse waters. An alternating current is passed between two electrodes in contact with a solution. The transmitter converts the resultant potential difference into conductivity, concentration or specific resistance.

2-Electrode Contacting Cells (United States Pharmacopœia <645>)
Traceable calibration instruments for WFI and PW processes in the Pharmaceutical Industry.

4-Electrode Contacting Cells - Range 0 to 2,000 mS/cm
Auto-compensating technology for high level conductivity and concentration measurements in high-coating processes. Provides an alarm when cleaning is required.

Electrodeless sensors - Range 0 to 2,000 mS/cm
Non-contact technique providing immunity to polarisation, oily fouling and non-conducting coatings until gross fouling occurs. Measurement of the magnitude of induced AC into a closed loop of solution using two embedded toroids.

2-Electrode Contacting Cells - Range 0 to 10,000 µS/cm

AX410
AX411
AX416
TB82TE*
Input 1
µS/cm
µS/cm
µS/cm
Input 2
-
µS/cm
pH
Application
4-wire AC & 24V DC
4-wire AC & 24V DC
4-wire AC & 24V DC
Communications
4 to 20mA, Profibus DP
4 to 20mA, Profibus DP
4 to 20mA, Profibus DP
Mounting
Panel, Wall, Pipe
Panel, Wall, Pipe
Panel, Wall, Pipe
[
TB82TE*
TB84TE*
Input
µS/cm
µS/cm
Application
2-wire 24V DC
4-wire AC
Communications
4 to 20mA, HART, FF,
Profibus PA
4 to 20mA
Mounting
Panel, Wall, Pipe
Panel, Wall, Pipe
* Range 0 to 20,000 µS/cm
[
[
[
2-Electrode Contacting Cells - United States Pharmacopœia <645>
AX450
AX455
Input 1
µS/cm
µS/cm
Input 2
-
µS/cm
Application
4-wire AC
4-wire AC
Communications
4 to 20mA, Profibus DP
4 to 20mA, Profibus DP
Mounting
Panel, Wall, Pipe
Panel, Wall, Pipe
[[
[
[
4-Electrode Contacting Cells - Range 0 to 2,000 mS/cm
AX430
AX436
TB82EC**
TB84EC**
Method
4-electrode
4-electrode
4-electrode
4-electrode
Input 1
µS/cm
µS/cm
mS/cm
mS/cm
Input 2
-
pH
-
-
Application
4-wire AC
4-wire AC
2-wire 24V DC
4-wire AC
Communications
4 to 20mA, Profibus
4 to 20mA, Profibus
4 to 20mA, HART,
FF, Profibus PA
4 to 20mA
Mounting
Panel, Wall, Pipe
Panel, Wall, Pipe
Panel, Wall, Pipe
Panel, Wall, Pipe
** Range 0 to 400 mS/cm
[[
[
[

Electrodeless sensors - Range 0 to 2,000 mS/cm
TB82TC**
TB84TC**
Method
Electrodeless
Electrodeless
Input 1
mS/cm
mS/cm
Input 2
-
-
Application
2-wire 24V DC
4-wire AC
Communications
4 to 20mA, HART,
FF. Profibus PA
4 to 20mA
Mounting
Panel, Wall, Pipe
Panel, Wall, Pipe

    •   Cancel
      • Twitter
      • Facebook
      • LinkedIn
      • Weibo
      • Print
      • Email
    •   Cancel

    Your preferences:

    ABB contact for United States of America

    Downloads

    View related documents and downloads for:
    seitp330 270180329ad03d7cc1256cd9003aa5bc